Fluoroscopy Screen Parameters in the Cath Lab: A Complete Guide for Cardiologists and Electrophysiologists
Have a look on his example fluoroscopy screen of Philips cath lab/fluoroscopy monitor during coronary angiography (CAG). Here's what the displayed parameters mean:
Display Meaning
CAG 15 f/s Coronary angiography at 15 frames/second (cine acquisition). Higher frame rate = smoother motion but higher radiation.
LAO/RAO: 0° No Left Anterior Oblique or Right Anterior Oblique angulation. Straight AP view.
CRA/CAU: 0° No cranial or caudal tilt. AP projection.
SID: 95 cm Source-to-Image Distance = distance from X-ray tube focal spot to detector.
8" Field of view (FOV) is 8 inches. Smaller FOV provides magnification and better spatial resolution.
0° (rotation icon) Detector rotation is neutral.
89 cm Table height (approximately detector/patient positioning parameter).
38 cm Detector position or source-to-patient distance parameter (machine dependent).
1° Slight detector/head tilt.
0% Image processing/roadmap or contrast enhancement parameter (vendor-specific).
Eye icon + L LIH (Last Image Hold) is active. The displayed image is the last acquired fluoroscopic frame.
5.0 min Total fluoroscopy time during the procedure so far.
Exposure parameters (bottom)
Fluoroscopy (F)
80 kV = Tube voltage
20 mA = Tube current
3.3 ms = Pulse width/exposure time per pulse
Radiography/Cine (R)
73 kV
173 mA
3.2 ms
Notice that cine uses much higher tube current (173 mA vs 20 mA), which is why cine images have better quality but deliver a higher radiation dose.
Orientation markers
LIH = Last Image Hold
L = Left side of the patient
Warning message
> "Init setting of collimator required. Call service."
This indicates the system believes the collimator requires initialization or calibration. It is generally a service/maintenance message and should be evaluated by the equipment engineer if it persists. It does not necessarily mean the current procedure cannot continue, but the collimator may not be functioning optimally.
.....
Understanding Fluoroscopy Screen Parameters in the Cath Lab: A Complete Guide for Cardiologists and Electrophysiologists
Whether you are performing coronary angiography, pacemaker implantation, CRT, ICD procedures, or catheter ablation, understanding every parameter displayed on the fluoroscopy monitor is essential. These values directly affect image quality, radiation exposure, procedural efficiency, and patient safety.
This guide explains the common fluoroscopy screen parameters seen on modern cath lab systems.
---
1. LIH (Last Image Hold)
Definition
Last Image Hold (LIH) is the last fluoroscopic image displayed after you release the fluoroscopy pedal.
Purpose
Reduces radiation exposure
Allows catheter position review without additional X-rays
Helps plan the next catheter movement
Improves procedural efficiency
Key Point
No new radiation is generated while reviewing an LIH image.
---
2. CAG 15 f/s
Meaning
Coronary Angiography recorded at 15 frames per second.
The machine acquires 15 individual images every second during cine recording.
Common frame rates include:
7.5 fps
10 fps
15 fps
30 fps (rare today)
Higher frame rate
Advantages
Better temporal resolution
Smoother visualization of moving catheters
Improved coronary imaging
Disadvantages
Increased radiation dose
Larger storage requirements
Current recommendations
Routine coronary angiography:
10–15 fps
Electrophysiology procedures:
7.5 fps is usually sufficient
---
3. LAO / RAO
Left Anterior Oblique / Right Anterior Oblique Angulation
These represent horizontal rotation of the C-arm around the patient.
Examples
LAO 45°
The X-ray tube rotates toward the patient's left.
RAO 30°
The tube rotates toward the patient's right.
Why important?
Different angulations separate overlapping cardiac structures.
Examples
Coronary angiography
LAO Cranial → Left main bifurcation
RAO Caudal → Circumflex artery
LAO Caudal → Spider view
Electrophysiology
LAO 45° differentiates septal vs lateral structures
RAO 30° assesses anterior-posterior catheter position
---
4. Cranial (CRA) / Caudal (CAU)
These indicate vertical tilt of the C-arm.
Cranial
The X-ray beam points toward the patient's head.
Caudal
The beam points toward the feet.
Uses
Coronary imaging
Cranial views elongate LAD
Caudal views visualize circumflex and left main
Electrophysiology
Useful for
Coronary sinus lead positioning
His bundle localization
Left bundle branch pacing
Septal visualization
---
5. SID (Source-to-Image Distance)
Definition
Distance between the X-ray tube focal spot and the detector.
Example
95 cm
Effects
Increasing SID
Advantages
Reduced geometric magnification
Better image sharpness
Disadvantages
May require higher radiation output
Modern systems automatically compensate.
---
6. Field of View (FOV)
Example
8-inch field
Field of View determines the size of anatomy displayed.
Common sizes
6 inch
8 inch
10 inch
12 inch
Smaller field
Advantages
Higher magnification
Better spatial resolution
Disadvantages
Higher patient radiation
Smaller visible area
Larger field
Advantages
Wider anatomy
Lower magnification
---
7. Fluoroscopy Time
Example
5.0 minutes
This is the cumulative duration during which fluoroscopy has been active.
Important
Fluoroscopy time alone does not accurately represent radiation dose.
Two procedures with identical fluoroscopy times may deliver very different doses depending on:
Patient size
Pulse rate
Frame rate
Magnification
Angulation
Dose settings
---
8. Tube Voltage (kV)
Example
80 kV
kV determines X-ray energy.
Higher kV
Advantages
Better penetration
Lower image noise
Disadvantages
Reduced image contrast
Lower kV
Advantages
Better contrast
Disadvantages
Less penetration
Higher patient dose in large individuals
---
9. Tube Current (mA)
Example
20 mA (Fluoroscopy)
173 mA (Cine)
mA determines the number of X-ray photons produced.
Higher mA
Advantages
Better image quality
Less noise
Disadvantages
Increased radiation dose
Notice that cine acquisition uses much higher mA than fluoroscopy because high-quality images are required for permanent recording.
---
10. Pulse Width (ms)
Example
3.3 milliseconds
Pulse width is the duration of each X-ray pulse.
Short pulse width
Advantages
Less motion blur
Sharper catheter visualization
Long pulse width
Advantages
More photons
Less image noise
Disadvantages
More motion blur
---
11. Fluoroscopy (F)
Displayed as
F
Represents live, low-dose continuous or pulsed imaging.
Characteristics
Lowest radiation mode
Used for catheter manipulation
Not intended for permanent storage
---
12. Cine (R)
Displayed as
R
Represents recorded radiographic acquisition.
Characteristics
High image quality
Higher radiation dose
Permanently stored
Used for coronary angiography and documentation
Radiation during cine may be 5–10 times higher than during fluoroscopy, depending on system settings.
---
13. Detector Rotation
Some systems display detector rotational angle.
This indicates detector orientation relative to the patient.
Used for
Optimizing visualization
Reducing overlap
Improving procedural ergonomics
---
14. Table Height
Example
89 cm
Represents patient table position.
Proper table height
Improves operator ergonomics
Reduces radiation scatter
Optimizes image quality
---
15. Detector Position
Some systems display detector distance from the patient.
General rule
Keep the detector as close to the patient as possible.
Benefits
Lower radiation dose
Better image quality
Less magnification distortion
---
16. Dose Area Product (DAP/KAP)
Although not shown in your image, this is one of the most important radiation parameters.
Definition
DAP (or Kerma Area Product, KAP) is the total X-ray energy delivered multiplied by the irradiated area.
Unit
Gy·cm²
Uses
Estimates stochastic radiation risk
Tracks cumulative radiation exposure
Required for quality assurance
---
17. Air Kerma (AK)
Usually displayed separately.
Definition
Radiation delivered to a reference point near the patient's skin.
Unit
Gray (Gy)
Importance
Air Kerma correlates with the risk of skin injury.
Clinical thresholds
≥2 Gy: Increased awareness
≥5 Gy: Follow-up recommended
≥10 Gy: Significant risk of deterministic skin injury
---
18. Collimator Warning
Example
"Init setting of collimator required. Call service."
Meaning
The collimator requires initialization or calibration.
Possible causes
System startup issue
Mechanical calibration error
Service requirement
If persistent, biomedical engineering or the equipment vendor should evaluate the system.
---
Radiation Reduction Tips for Every Procedure
Use LIH instead of repeated fluoroscopy.
Prefer 7.5 fps for most electrophysiology procedures.
Keep the detector close to the patient.
Maximize the distance between the X-ray tube and the patient.
Use tight collimation.
Avoid unnecessary cine acquisitions.
Use lower magnification whenever possible.
Minimize steep LAO/cranial projections, which often require higher radiation output.
Wear appropriate shielding and monitor operator radiation exposure.
---
Key Takeaway
Every number displayed on the fluoroscopy monitor has a practical impact on image quality, radiation dose, and procedural success. A clear understanding of parameters such as LIH, frame rate, LAO/RAO, cranial/caudal angulation, SID, field of view, fluoroscopy time, kV, mA, pulse width, DAP, and Air Kerma enables cardiologists and electrophysiologists to optimize imaging while adhering to the ALARA (As Low As Reasonably Achievable) principle for radiation safety.
For more cardiology and electrophysiology educational content, visit: drmusmanjaved.com

Comments
Post a Comment
Drop your thoughts here, we would love to hear from you